Remote Control Commands
R&S
®
NRPxxA(N)
89
User Manual 1177.6017.02 ─ 05
Example:
BUFF:CLE
Usage:
Event
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?
Queries the number of results that are currently stored in the result buffer. Available in
continuous average mode.
Example:
BUFF:COUN?
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:DATA?
Queries the results of the continuous average result buffer and returns them even if the
buffer is not full.
In contrast the
FETCh?
command returns a result only if the buffer is full.
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
<count>
Sets the size of the result buffer in continuous average mode.
You can enable the buffer using
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:
.
Parameters:
<count>
Range:
1 to 8192
*RST:
1
Example:
BUFF:SIZE 1
See
Chapter 10.3, "Performing a Buffered Continuous Average
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe
<state>
Enables or disables the buffered continuous average mode. If the buffer mode is
enabled, all results generated by trigger events are collected in the sensor until the buf-
fer is filled.
You can set the size of the buffer with
.
Parameters:
<state>
ON | OFF
*RST:
OFF
Example:
BUFF:STAT OFF
Configuring the Measurement Modes