Operating Concepts
R&S
®
NRPxxA(N)
37
User Manual 1177.6017.02 ─ 05
Required equipment
●
R&S NRPxxA(N) power sensor
●
R&S NRP
‑
ZKU cable or an R&S NRP
‑
Z5 sensor hub and an R&S NRP
‑
ZK6 cable
to connect the sensor to the computer
●
Computer with installed:
–
R&S NRP Toolkit V
4.17 or higher
–
R&S Power Viewer version 9.2 or higher (refer to the operating manual of the
R&S Power Viewer for a description of the installation process)
If you want to use an android device like a tablet or a smartphone, use the R&S Power
Viewer Mobile. For details, see
Chapter 5.6, "R&S Power Viewer Mobile"
Setup
Figure 5-3: Setup with the R&S Power Viewer
1 = Signal source
2 = R&S NRPxxA(N) power sensor
3 = Host interface connector
4 = R&S NRP
‑
ZKU cable
5 = USB connector
6 = Computer with installed R&S Power Viewer
Incorrectly connecting/disconnecting the R&S NRPxxA(N) power sensors can damage
the power sensors or lead to erroneous results.
Ensure that you connect/disconnect your power sensor as described in
Starting a measurement
1. Connect the cables as shown in
For a detailed description, refer to
Chapter 3.5.1.1, "Simple USB Connection"
2. Start the R&S Power Viewer.
3. Execute zeroing:
R&S Power Viewer