Remote Control Commands
R&S
®
NRPxxA(N)
94
User Manual 1177.6017.02 ─ 05
[SENSe<Sensor>:]AVERage[:STATe]
<state>
Available in continuous average mode.
Enables or disables the averaging filter.
Parameters:
<state>
*RST:
ON
Manual operation:
See
9.8.2
Setting the Frequency
The frequency of the signal to be measured is not automatically determined. For ach-
ieving better accuracy, the carrier frequency of the applied signal must be set.
[SENSe<Sensor>:]FREQuency
<frequency>
Transfers the carrier frequency of the RF signal to be measured. This frequency is
used for the frequency-response correction of the measurement result.
The center frequency is set for broadband signals, e.g. spread-spectrum signals, multi-
carrier signals.
Parameters:
<frequency>
Range:
0.0 to 110.0e9
*RST:
50.0e6
Default unit: Frequency
Example:
FREQ 10000
Manual operation:
See
9.8.3
Configuring Corrections
It is possible to set some parameters that compensate for a change of the measured
signal due to fixed external influences.
........................................................................................... 94
.......................................................................................... 96
............................................................................................ 98
................................................................................ 100
9.8.3.1
Duty Cycle Corrections
The duty cycle is the percentage of one period during which the signal is active, when
pulse-modulated signals are corrected. The duty cycle is only evaluated in the Continu-
ous Average mode.
Configuring Basic Measurement Parameters