Remote Control Commands
R&S
®
NRPxxA(N)
85
User Manual 1177.6017.02 ─ 05
FORMat[:DATA]
[<data,length>, <length>]
Specifies how the R&S NRPxxA(N) sends the numeric data to the controlling host/
computer.
Parameters:
<data,length>
<REAL,32 | 64>
REAL
Block of binary values, 32-bit or 64-bit each; so-called "SCPI
definite length block"
32 | 64
32-bit or 64-bit
If you omit the length, the R&S NRPxxA(N) sets the last used
length.
Example for
REAL,32
format:
#6768000....<binary float values>....
Example for
REAL,64
format:
#71536000....<binary float values>....
<data[,length]>
<ASCii[,0 to 12]>
ASCii
List of comma separated, readable values.
[,0 to 12]
Defines the number of decimal places.
The reset value 0 does not restrict the number of decimal pla-
ces.
Example for
ASCii,4
format:
1.2938e-06, -4.7269e-11, ...
*RST:
ASCii,0
9.6.3
Retrieving Results
After performing the measurement, you can query the measurement results with a
command from the
FETCh
subsystem.
Remote commands:
FETCh<Sensor>:ARRay[:POWer][:AVG]?
.......................................................................... 85
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
........................................................................86
FETCh<Sensor>[:SCALar][:POWer]:TSLot?
....................................................................... 86
..............................................................................................86
FETCh<Sensor>:ARRay[:POWer][:AVG]?
Queries the last valid measurement result of a measurement with enabled data buffer
mode.
Usage:
Query only
Configuring and Retrieving Results