Power Sensor Tour
R&S
®
NRPxxA(N)
27
User Manual 1177.6017.02 ─ 05
Indication
State
Slow flashing white
Sanitizing in progress
Yellow
Wait for trigger state
Green
Measuring state
Turquoise blue
Zeroing is in progress
Slow flashing red
Static error
You can query the error type with
.
Fast flashing red
Critical static error
You can query the error type with
.
Note:
If this state occurs after a firmware update, the update was not
successful. Perform the firmware update again.
Chapter 12.3, "Problems during a Firmware Update"
4.5
LAN PoE Interface
Available only for LAN power sensor.
An RJ-45 connector is used to connect the Ethernet interface of the power sensors to a
Local Area Network (LAN).
Ethernet interface requires PoE (Power over Ethernet)
When using the Ethernet interface of the power sensors, the electrical power has to be
provided by Power over Ethernet (PoE). In this case, it is
not
possible to provide the
power supply via the USB connector instead.
Risk of sensor damage
Use only PoE power sourcing equipment (PSE) according to IEEE standards 802.3af
or IEEE 802.3at.
Otherwise your power sensor can get damaged.
LAN reset button
The LAN reset button is used for resetting the Ethernet connection parameters of the
power sensor to their default values.
Power over Ethernet status LED
Available only for LAN power sensor.
The power status LED shows whether the sensor is correctly powered over PoE or not.
LAN PoE Interface