Remote Control Commands
R&S
®
NRPxxA(N)
122
User Manual 1177.6017.02 ─ 05
SYSTem:RESTart
Restarts the firmware of the R&S NRPxxA(N).
Usage:
Event
SYSTem:RUTime
<update_time>
Effective only in the NRP legacy mode.
Sets the result update time. That is the maximum rate in which the power sensor can
output measurement results. Relevant only in continuous measurement mode
(
ON).
Parameters:
<update_time>
Range:
0.0 to 10.0
*RST:
0.1
Default unit: Seconds
SYSTem:SERRor?
Queries the next static error, if available. Static errors are more severe than normal
error conditions, which can be queried using
Normal errors result from, for example, unknown commands or syntax errors and gen-
erally affect a single parameter or setting. Static errors, as a rule, prevent the execution
of normal measurements.
Usage:
Query only
SYSTem:SERRor:LIST:ALL?
Returns a list of all static errors that have occurred but have already been resolved.
For example, an overload of a short duration.
Example:
SYST:SERR:LIST:ALL?
Response:
0,"reported at uptime:2942; notice;
auto-averaging exceeded maximum time;
Notification",0,"removed at uptime:2944;
notice; auto-averaging exceeded maximum time;
Notification"
.
Usage:
Query only
SYSTem:SERRor:LIST[:NEXT]?
Queries the list of all static errors that have occurred but have already been resolved
for the oldest entry and removes it from the queue. The response consists of an error
number and a short description of the error.
Configuring the System