Operating Concepts
R&S
®
NRPxxA(N)
38
User Manual 1177.6017.02 ─ 05
Note:
Turn off all measurement power signals before zeroing. An active measure-
ment signal during zeroing causes an error.
a) Switch off the measurement signal.
b) Select "Sensor > Zero (Signal off) ".
4. Switch on the test signal of the signal source.
5. For a continuous average measurement, select "Measurement > Continuous".
The "Continuous" measurement window appears. It shows the measurement
results numerically and some parameters that can be configured.
6. To start the measurement press "Measurement > Start".
The measurement result is shown in the "Continuous" measurement window.
For a detailed description of how to measure in this setup, refer to the operating man-
ual of your R&S Power Viewer. The manual is installed automatically during the instal-
lation of the R&S Power Viewer.
R&S Power Viewer