Troubleshooting
R&S
®
NRPxxA(N)
157
User Manual 1177.6017.02 ─ 05
12
Troubleshooting
............................................................................... 157
.............................................................................................157
Problems during a Firmware Update
.................................................................... 158
Cannot Establish a LAN Connection
.....................................................................158
12.1
Displaying Status Information
Status information is available in several ways.
Status LED of the R&S NRPxxA(N)
The position of the status LED is indicated in
Chapter 4, "Power Sensor Tour"
The meaning of the different colors and blinking frequencies is explained in
Title bar of the web user interface (LAN power sensors)
Only available for power sensors with networking capabilities, the R&S
NRP LAN
power sensors.
The position of the status icon is indicated in
12.2
Performing a Selftest
The selftest gives you detailed information that you can use for troubleshooting.
Do not apply a signal to the sensor while the selftest is running. If the selftest is carried
out with a signal being present, error messages can erroneously be output for the test
steps
Offset Voltages
and/or
Noise Voltages
.
Using remote control
.
For each test step, PASS or FAIL is listed.
Using the web user interface (LAN power sensors)
1. In the navigation pane of the main dialog, select "Sensor".
2. Under "Diagnostics", click "Selftest".
Performing a Selftest