Remote Control Commands
R&S
®
NRPxxA(N)
82
User Manual 1177.6017.02 ─ 05
If the sensor triggers itself, the trigger source of the sensor must be assigned to the
other external port, as shown in the examples.
Parameters:
<master_port>
EXT1 | EXTernal1 | EXT2 | EXTernal2
*RST:
EXT1
Example:
TRIG:MAST:PORT EXT1
TRIG:SOUR EXT2
TRIG:MAST:STAT ON
Example:
TRIG:MAST:PORT EXT2
TRIG:SOUR EXT1
TRIG:MAST:STAT ON
TRIGger:MASTer:STATe
<state>
Enables or disables the trigger master mode of the sensor. In this state, the power sen-
sor can output a digital trigger signal in sync with its own trigger event.
If enabled, select the output port for the trigger signal using
.
Typically, the trigger master uses its internal trigger source. But you can also trigger the
trigger master externally, because the power sensor has got two external trigger con-
nectors. If you trigger the trigger master externally, use
EXTernal1
as external trigger
input port (trigger source) and
EXTernal2
as trigger master output port or vice versa.
Parameters:
<state>
*RST:
OFF
TRIGger:SLOPe
<slope>
Effective only if
INTernal
or
EXTernal
.
Determines which edge of the envelope power, with internal triggering, or increasing
voltage, with external triggering, is used for triggering.
Parameters:
<slope>
POSitive | NEGative
POSitive
Rising edge
NEGative
Falling edge
*RST:
POSitive
Manual operation:
See
TRIGger:SOURce
<source>
Selects the source for the trigger event detector.
Controlling the Measurement