Remote Control Commands
R&S
®
NRPxxA(N)
99
User Manual 1177.6017.02 ─ 05
Sensor
Source
S
21
S
12
S
22
S
11
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (−70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
Figure 9-4: Correction of interactions between the power sensor, the signal source, and the S-param-
eter device
Remote commands:
[SENSe<Sensor>:]SGAMma:CORRection:STATe
................................................................99
[SENSe<Sensor>:]SGAMma:MAGNitude
........................................................................... 99
................................................................................. 99
[SENSe<Sensor>:]SGAMma:CORRection:STATe
<state>
Enables or disables the use of the complex reflection coefficient to correct the interac-
tions of power sensor and signal source.
Parameters:
<state>
*RST:
OFF
Manual operation:
See
[SENSe<Sensor>:]SGAMma:MAGNitude
<magnitude>
Sets the magnitude of the complex reflection coefficient of the source,
Γ
source
.
Parameters:
<magnitude>
0.0
≙
ideal matched source
1.0
≙
total reflection
Range:
0.0 to 1.0
*RST:
0.0
Manual operation:
See
[SENSe<Sensor>:]SGAMma:PHASe
<phase>
Sets the phase angle of the complex reflection coefficient of the source,
Γ
source
.
Parameters:
<phase>
Range:
-360.0 to 360.0
*RST:
0.0
Manual operation:
See
Configuring Basic Measurement Parameters