Remote Control Basics
R&S
®
NRPxxA(N)
148
User Manual 1177.6017.02 ─ 05
The parallel poll enable register can be set and read by
.
11.3.5
Device Status Register
1
0
2
3
4
5
6
7
9
8
10
11
12
13
14
15
+
Device
Status
Sensor Error Summary
Sensor Error
Sensor Error
Sensor Error
Sensor Error
0
0
Legacy Locked
Reference-PLL Locked
0
0
0
0
0
0
0
Querying the register:
●
●
Querying the static errors:
●
Table 11-5: Meaning of bits used in the device status register
Bit No.
Meaning
0
Sum of SERR bits
The sum/combination of SERR bits 1 to 4.
1
SERR measurement not possible
Static error (SERR) exists; certain parameter settings could lead to a situation where subse-
quent measurements are not possible
2
SERR erroneous results
Static error exists; the measurement result is possibly incorrect.
3
SERR warning
A static error exists therefore the Status LED of the power sensor is blinking slowly in red.
4
SERR critical
A critical static error exists therefore the Status LED of the power sensor is blinking fast in
red.
Status Reporting System