CDMA2000 Measurements
R&S
®
CMW-KM8xx/-KS8xx
764
User Manual 1173.9511.02 ─ 12
5.3.9 OLTR: Measurement Control
shares the routing, analyzer and trigger settings with the
Multi Evaluation measurement. The OLTR-specific "Measurement Control" parameters
can be set in the configuration dialog of the OLTR measurement.
Fig. 5-27: OLTR configuration
................................................................................................ 764
.......................................................................................764
................................................................................................ 765
Measurement Timeout
Defines a timeout for the OLTR measurement (in remote operation).
Remote command:
CONFigure:CDMA:MEAS<i>:OLTR:TOUT
Number of Meas Sequences
Sets the number of measurement sequences within a single OLTR measurement.
Each sequence consists of a power UP or power DOWN step, followed by a power
step in the opposite direction (see "Power Step Direction" below).
Remote command:
CONFigure:CDMA:MEAS<i>:OLTR:SEQuence
Power Step
Defines the size of the power steps, i.e. the increases and decreases in the total BSS
power in the forward link during the OLTR measurement.
Remote command:
CONFigure:CDMA:MEAS<i>:OLTR:PSTep
Power Step Direction
Defines the direction of the first power step in the forward link within an OLTR mea-
surement. For each subsequent power step the direction is toggled.
Remote command:
CONFigure:CDMA:MEAS<i>:OLTR:PSTep:DIRection