CDMA2000 Signaling
R&S
®
CMW-KM8xx/-KS8xx
596
User Manual 1173.9511.02 ─ 12
●
27 (No Sensor Connected)
:
The GPRF External Power Sensor measurement could not be started due to miss-
ing power sensor.
●
30 (File not Found)
:
The specified file could not be found.
●
40 (ARB File CRC Error)
:
The ARB file CRC check failed. The ARB file is corrupt and not reliable.
●
42 (ARB Header Tag Invalid)
:
The ARB file selected in the GPRF generator contains an invalid header tag.
●
43 (ARB Segment Overflow)
:
The number of segments in the multi-segment ARB file is higher than the allowed
maximum.
●
44 (ARB File not Found)
:
The selected ARB file could not be found.
●
45 (ARB Memory Overflow)
:
The ARB file length is greater than the available memory.
●
46 (ARB Sample Rate out of Range)
:
The clock rate of the ARB file is either too high or too low.
●
47 (ARB Cycles out of Range)
:
The repetition mode equals "Single Shot" and the playback length is greater than
40 s. Please reduce the playback length or set the repetition mode to "Continuous".
<Length> = (<Cycles> * <Samples> + <Additional Samples>) / <Clock Rate>
●
50 (Startup Error)
:
The Data Application Unit (DAU), a DAU service or a DAU measurement could not
be started. Please execute a DAU selftest.
●
51 (No Reply)
:
The DAU has received no response, for example for a ping request.
●
52 (Connection Error)
:
The DAU could not establish a connection to internal components. Please restart
the instrument.
●
53 (Configuration Error)
:
The current DAU configuration by the user is incomplete or wrong and could not be
applied. Check especially the IP address configuration.
●
54 (Filesystem Error)
:
The hard disk of the DAU is full or corrupt. Please execute a DAU selftest.
●
60 (Invalid RF-Connector Setting)
The individual segments of a list mode measurement with R&S CMWS use differ-
ent connector benches. This is not allowed. All segments must use the same
bench.
Check the "Info" dialog for the relevant segment numbers.
●
101 (Firmware Error)
:
Indicates a firmware or software error. If you encounter this error for the first time,
restart the instrument.
Command Reference