CDMA2000 Signaling
R&S
®
CMW-KM8xx/-KS8xx
623
User Manual 1173.9511.02 ─ 12
Firmware/Software:
V3.2.80
Options:
R&S CMW-KE100 and R&S CMW-KE800
Manual operation:
See
ROUTe:CDMA:SIGN<i>:SCENario:SCFading[:EXTernal]
<RXConnector>,
<RXConverter>, <TXConnector>, <TXConverter>, <IQConnector>
Activates the "Standard Cell Fading: External" scenario and selects the signal paths.
For possible connector and converter values, see
chapter 2.5.1.3, "Values for Signal
Parameters:
<RXConnector>
RF connector for the input path
<RXConverter>
RX module for the input path
<TXConnector>
RF connector for the output path
<TXConverter>
TX module for the output path
<IQConnector>
DIG IQ OUT connector for external fading of the output path
Example:
See
Specifying General Signaling Settings
Firmware/Software:
V3.2.80
Manual operation:
See
CONFigure:CDMA:SIGN<i>:RFSettings:EATTenuation
<RFInputExtAtt>,
<RFOutputExtAtt>
Defines the external attenuations (or gain, if the value is negative), to be applied to the
selected RF input and output connectors.
Parameters:
<RFInputExtAtt>
Range:
-50 dB to 90 dB
*RST:
0 dB
Default unit: dB
<RFOutputExtAtt>
Range:
-50 dB to 90 dB
*RST:
0 dB
Default unit: dB
Example:
See
Specifying General Signaling Settings
Firmware/Software:
V1.0.15.0
Manual operation:
See
CONFigure:CDMA:SIGN<i>:IQIN:PATH<n>
<PEP>, <Level>
Specifies properties of the baseband signal at the I/Q input.
Command Reference