- 38 -
The next measurement is possible when the indication gets lower than the value
set for <AUTO THRESHOLD> parameter (next measurement does not require
zero value).
For automatic operation it is necessary to adjust the threshold value. The measurement is
sent to a computer or a printer when mass indication gets below the set net value of
<AUTO
THRESHOLD>
parameter.
<AUTO THRESHOLD>
parameter refers to the following functions: automatic tare,
automatic operation, autoprint with interval.
11.15. Minimum Sample Weight
Weighing mode comprises <Minimum sample weight> function. In order to use this function
it is necessary to enter minimum sample weight (MSW) value and tare values for which the
MSW value is to be obligatory. For standard PM C32 series the values are zero.
Only authorized RADWAG employee or operator with Administrator permissions level,
providing that balance factory settings enable this, can carry out procedure aiming to
determine minimum sample weight and next enter the respective data.
If you want to use this function, and your balance menu contains no minimum sample weight
data, ask the nearest RADWAG office for help.
Using reference sample weights, the authorized RADWAG employee determines minimum
sample weight for specified tare containers. This operation is carried out at the place of use
and in accordance with requirements of the applied quality system. The obtained value is
entered into the weighing instrument program.
The weighing instrument program enables defining tare value with assigned minimum
sample weight value.
<Minimum sample weight> function guarantees that the results of weighing operation are
comprised within the set tolerance, in accordance with applied quality management system
of particular company.
Caution: Function valid for weighing mode exclusively.
Available options:
•
MODE
None
– minimum sample weight function off.
Block
– select to enable display of respective pictograms informing you about the
measured mass (whether it is lower than the MSW value or higher than the MSW
value); with this option on, the program makes it impossible to accept the measurement
that is lower than the MSW value.
Warn
– select to enable display of respective pictograms informing you about the
measured mass (whether it is lower than the MSW value or higher than the MSW
value). You can accept the measurement that is lower than the MSW value.
•
Tare
– maximum tare value, for which the minimum sample weight is obligatory
(analyse the below examples).
•
Minimum mass –
value of the minimum sample weight determined for a particular
weighing device on site with use of respective method.
Example 1:
No.
Tare value
Minimum
Sample
Weight
Operation
2
1000.00 g
1.00 g
Minimum sample weight refers to all net weight values
when weighing is carried out using a tare container of
mass 0.01 g – 999.99 g, inclusive (<TARE> button used).