- 57 -
•
Put the container (intended for the reference sample) on a weighing pan and tare it.
•
Press
button (Determine part mass), <Reference sample quantity> edit box
featuring an on-screen numeric keyboard is displayed.
•
Enter respective value and press
key to confirm changes.
•
Message specifying number of parts to be loaded is displayed:
<Number of parts to
be loaded: 18>
.
•
Load the weighing pan with declared amount of parts. When the indication is stable (
pictogram is displayed), press
key to confirm the mass.
•
Single part mass is calculated automatically, <Parts counting> working mode is run and
the weighing device displays quantity of single parts (
pcs
).
Caution: Remember:
•
Total weight value of all parts loaded onto the weighing pan cannot be greater than
the max capacity value.
•
Total weight value of all parts loaded onto the weighing pan cannot be lower than the
value determined by <Minimum reference sample mass> parameter. Unless this
condition is met, the weighing device displays a message: <Sample mass too low>.