18.3. Making Use of Checkweighing Thresholds
There are two methods for making use of the checkweighing thresholds:
−
selecting product <
> with Min and Max threshold declared,
−
entering numeric value for the thresholds using <
> or <
> or <
> button, in
this case thresholds do not refer to any product.
Procedure 1
– selecting product from Products Database:
•
Press
button (product selection), list of products is displayed.
•
Select the product that is to be weighed.
•
The balance displays threshold values under the bar graph within the INFO area. Bar
graph colour corresponds to the current status of mass:
o
yellow: mass value lower than Min value,
o
green: mass value higher than MIN value and lower than MAX value,
o
red: mass value higher than MAX value.
Procedure 2
– entering checkweighing thresholds manually:
•
Press
button (Checkweighing thresholds).
•
Window with an on-screen numeric keypad is displayed, using it enter Min Threshold
value.
•
Enter the value, press
key.
•
Window with an on-screen numeric keyboard is displayed, using it enter Max
Threshold value.
•
Enter the value, press
key.
•
Home screen of the Checkweighing mode along with the entered threshold values is
displayed.
Caution: Max threshold value has to be greater than Min threshold value.
In order to modify threshold values, the following buttons can be used:
•
- press to modify Min Threshold value, do it using an on-screen numeric keypad.
Enter the value, press
key. Home screen of the Checkweighing mode is
displayed automatically.