- 37 -
•
When the weighing device detects weight value lower than gross weight value set
in <
AUTO THRESHOLD
> parameter, the entered tare value gets automatically
deleted (
Net
pictogram is no longer displayed).
•
Load the weighing pan with packaging of another product, upon indication
stabilization, packaging mass is automatically tared (
Net
pictogram is displayed in
the upper part of the display).
•
Put/pour/dispense product that is to be packed into the packaging.
For correct operation of the weighing instrument with AUTOTARE function, it is necessary to
adjust the threshold value.
<AUTO THRESHOLD>
parameter refers to the following functions:
- autotare,
- automatic printout.
No automatic taring takes place as long as the gross weight value stays within the range set
in
<AUTO THRESHOLD>
parameter.
11.14. Print Mode
Function designed to enable print mode setting.
Available values:
•
<WHEN STABLE> - for this value stable measurement result, along with the settings
for <GLP PRINTOUT> parameter, is sent to the printer port. Upon pressing
key at the moment when the result is unstable (no
pictogram
displayed), the program first waits for the stability condition to be met, next sends
the measurement result to the port.
•
<EACH> - every single pressing of
key results with sending the
measurement indication to the printer port along with the settings for <GLP
PRINTOUT> parameter. Every single indication is sent (stable and unstable). In
case of unstable indication, <?> sign is displayed in front of the weighing result
frame.
Function available for non-verified balances exclusively.
•
<AUTO> - select this value to enable automatic printout. When selecting <AUTO>
value remember to set <AUTOPRINT THRESHOLD> in accordance with your
requirements.
•
<AUTO + INTERVAL> - select this value to start automatic print and record of
indications in WEIGHINGS and ALIBI databases, carried out in a cyclic manner in a
specified time intervals. The interval is set in [min] in <INTERVAL> parameter.
Interval range is 1 - 9999 min.
CAUTION:
Each weighing result is printed and recorded (stable and unstable for a non-verified
balance, stable for a verified balance).
Automatic operation with interval starts at the moment of switching the function on.
The first stable weighing result of value greater than AUTO THRESHOLD value is
printed and recorded as the first measurement. The following measurements are
printed with frequency set in INTERVAL parameter. To stop the automatic operation
with interval, switch the option off.
Upon switching the function of auto print with interval on, PRINT button becomes
inoperative (no indication is printed when pressed).
Automatic operation procedure:
Press
key to zero the indication (stability marker,
, and zero
pictogram,
, are displayed).
Load the weighing pan, the first stable measurement is sent to the printer port.
Unload the weighing pan.