In case of such settings, the minimum sample weight is
valid only when you weigh samples in a tare container of
weight value comprised within the above specified range. If
taring function IS NOT used, or if the tare container's
weight ranges between 1000.00 g - Max, then pictogram
informing on use of minimum sample weight goes blank.
Example 2 (for a weighing instrument with max weighing range 10kg):
No.
Tare value
Minimum
Sample
Weight
Operation
1
10000.00 g
1.50 g
Minimum sample weight refers to all net weight values
when weighing is carried out using a tare container of
mass value comprised within the whole weighing range
(<TARE> button used).
In case of such settings, the minimum sample weight is
valid only when you weigh samples in a tare container. If
taring function IS NOT used then pictogram informing on
use of minimum sample weight goes blank.
Example 3:
No.
Tare value
Minimum
Sample
Weight
Operation
1
0.00 g
2.00 g
Minimum sample weight refers to all net weight values
when weighing is carried out without use of a tare
container (<TARE> button NOT used).
In case of such settings, the minimum sample weight is
valid only when you weigh samples without use of tare
containers. If taring function IS used then pictogram
informing on use of minimum sample weight goes blank.
You can preview the entered data, but you are not allowed to edit it.
Weighing with use of MINIMUM SAMPLE WEIGHT function
If in the course of weighing you want to find out whether a particular measurement value is
higher than the MSW value, specified for a given weighing range, then <Minimum sample
weight> function must be on. To turn the function on go to weighing mode settings.
Procedure (Administrator exclusively):
1. Enter weighing mode settings.
2. Enter <Minimum sample weight> parameter group.
3. Enter <Mode> parameter.
4. Settings box is displayed, select respective option:
Block
– select to enable display of respective pictograms informing you about the
measured mass (whether it is lower than the MSW value or higher than the MSW
value); with this option on, the program makes it impossible to accept the
measurement that is lower than the MSW value.
Warn
– select to enable display of respective pictograms informing you about the
measured mass (whether it is lower than the MSW value or higher than the MSW
value). You can accept the measurement that is lower than the MSW value.
5. Go to the home screen.
6. Section presenting the weighing result comprises additional pictogram providing
supplementary information. The pictogram changes in the course of weighing
operation, it informs you on the weighed sample mass; the information is referred to
the declared minimum sample weight value.