- 11 -
1.7. Supervision Over Metrological Parameters
Metrological parameters need to be checked in determined time intervals. Inspection
frequency depends on ambient conditions in which the weighing instrument is operated, kind
of carried out processes and adopted quality management system.
1.8.
User Manual Significance
Even if you are experienced and have already worked with this type of weighing instrument,
you are still obliged to read this user manual carefully prior switching the device on. This
user manual provides necessary operation-related information. Following the user manual
guidelines guarantees correct weighing instrument performance.
1.9.
Balance Operator Training
The weighing instrument must be utilized and supervised only by persons who have been
trained to operate such type of a weighing instrument.
2.
TRANSPORT AND STORAGE
2.1. Delivery Check
Upon delivery it is necessary to check the package and the device, make sure that your
package bears no signs of damage. Otherwise contact the manufacturer's representative.
2.2. Packaging
Keep all package elements should your device be transported in the future. Remember that
only original packaging can be used for shipping purposes. Prior packing, uncouple any
cables, remove any separable components (weighing pan, shields, inserts). The device
components shall be packed into an original packaging, thus being protected against
potential damage during transportation.
3.
UNPACKING AND INSTALLATION
3.1. Place of Use
•
Operate the device in a workroom free of vibrations and shakes, where there are no
air drafts nor dust.
•
Make sure that the ambient temperature ranges b10 °C ÷ +40 °C.
•
Make sure that the relative humidity is not higher than 80%.
•
Make sure that the temperature changes slowly in the course of operation.
•
Place the weighing instrument either on a robust-design table or on a wall bracket,
which is both distant from heat sources and insusceptible to vibrations.
•
Take special precaution while weighing magnetic objects, as part of the balance is a
strong magnet. Should such loads be weighed, use under-pan weighing option in
order to remove the weighed load from an area influenced by the magnet. The hook
for under-pan weighing is installed in the weighing instrument’s base.