- 117 -
Format:
US_x <CR><LF>
Response options:
US_ x_OK <CR><LF>
- command carried out, returns accessible units
US_E <CR><LF>
- error in-course of command execution, no parameter or
incorrect format
US_I <CR><LF>
- command understood but not accessible at this moment
x
- parameter, units symbols: g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht,
tola, msg, u1, u2, next
Caution:
If x=next the command swaps to another available unit on the list (it simulates
button
pressing).
Example:
Command:
US_mg<CR><LF>
- set „mg” unit
Response:
US_mg_OK<CR><LF>
- „mg” set as a current unit
Give Current Unit
Command overview:
Command returns current unit.
Format:
UG <CR><LF>
Response options:
UG_x_OK<CR><LF>
- command carried out, returns accessible units
UG_I <CR><LF>
- command understood but not accessible at this moment
x
- parameter, units symbols:
Example:
Command:
UG<CR><LF>
- return current unit
Response:
UG_ct_OK<CR><LF>
- currently set unit is „ct”
Activate Sound Signal
Command overview:
Command activates BEEP sound signal for a specified amount of time.
Format:
BP_TIME<CR><LF>
Response options:
BP_OK <CR><LF>
- command carried out, BEEP activated
BP_E” <CR><LF>
- no parameter or incorrect format
BP_I <CR><LF>
- command understood but not accessible at this moment
TIME
- parameter, decimal value specifying how long shall the sound last, parameter given
in [ms]. Recommended range <50
÷
5000>.
If value greater than the permissible high limit is given, then BEEP sound is operated for the
maximum permissible amount of time.
Example:
Command:
BP_350<CR><LF>
- activate BEEP for 350 ms
Response:
BP_OK<CR><LF>
- BEEP activated
Caution:
BEEP sound activated via BP command is inhibited if in-course of its activation the sound is
activated by means of other device: keypad, touch panel, proximity sensors.
Send All Implemented Commands
Содержание PM 10.C32
Страница 1: ...C32 Series Balances Precision balances of PM C32 series IMMU 23 04 11 17 EN www radwag com USER MANUAL...
Страница 7: ...7 1 GENERAL INFORMATION 1 1 Dimensions PM xx C32 d 0 01g PM xx C32 d 0 1g...
Страница 8: ...8 PM xx C32 d 0 5g d 1g Indicator...
Страница 125: ...125...
Страница 128: ...128...
Страница 129: ...129...