- 29 -
Log Out operation:
•
Go to home screen, press
key, wait for the operators database to open.
•
Select
<Log out>
option (
1st position on the operators list
).
•
The home screen is displayed.
Permissions Levels
There are 4 permissions levels: Administrator, Advanced Operator, Operator, None.
Access to operator-related parameters, databases and program functions is
conditioned by permissions level:
Permissions level
Available parameters and functions
None
Permission to edit operator-related parameters not granted.
Operator can neither accept the weighing result nor start the
following operations: entering reference sample mass and
determining reference sample quantity in working mode <Parts
counting>, entering reference sample mass and determining
reference sample in working mode <Percent weighing>.
Operator
Operator can edit parameters of <Readout/Filter> submenu, set
parameters of <Misc> submenu (except for <Date and time>,
<Permissions>, <Software update> parameters). Operator can
run and carry out all weighing operations. Operator can preview
<Databases> data, and define universal variables.
Advanced
operator
Operator can edit the following submenus: <Readout>;
<Working modes>; <Communication>; <Peripherals>; <Misc>
(operation of <Date and time>, <Permissions> and <Software
update> parameter excluded). Operator can run and carry out
all weighing operations.
Administrator
Operator can edit all operator-related parameters and all
databases, and use all functions.
11.
WEIGHING OPERATION
Load the weighing pan. Read the result when
stability marker is displayed .
In order to record/print the weighing result press <
> key:
•
for verified weighing instruments – only stable weighing result is recorded or printed
(stability marker
displayed),
•
for non-verified weighing instruments – stable or unstable weighing result is
recorded or printed (stability marker
not displayed
)
; Unstable weighing result is
marked with <
?
> on the printout, the question mark is printed next to the weight
value.
11.1. Good Weighing Practice
To assure long-term operation and correct mass measurements, follow the rules
presented below:
Start the balance with no load on the weighing pan (permissible value of load
on the weighing pan at start is ±10% of the maximum capacity).