- 119 -
Command:
A_1<CR><LF>
- turn autozero function on
Response:
A_OK<CR><LF>
- autozero function is on
AUTOZERO function operates until it is turned off by A 0 command.
Set Ambient Conditions State
Format:
EV_n <CR><LF>
Response options:
EV_OK <CR><LF>
- command carried out
EV_E <CR><LF>
- error in-course of command execution, no parameter or incorrect
format
EV_I <CR><LF>
- command understood but not accessible at this moment
n
- parameter, decimal value determining ambient conditions state
n
→
0 – unstable ambient conditions
1 – stable ambient conditions
Caution:
Command changes settings for a current working mode.
Example:
Command:
EV_1<CR><LF>
- set value ‘stable’ for ambient conditions option
Response:
EV_OK<CR><LF>
- set value ‘stable’ for ambient conditions option
<AMBIENT CONDITIONS> parameter is set to value <STABLE> until command EV 0 swaps
it to value <UNSTABLE>.
Give Set Ambient Conditions State
Command overview:
Command gives information about state of currently set ambient conditions.
Format:
EVG <CR><LF>
Response options:
EVG_x_OK<CR><LF>
- command carried out, response: set ambient conditions state
EVG_I <CR><LF>
- command understood but not accessible at this moment
x
- parameter, symbol of currently set ambient conditions state
Example:
Command:
EVG<CR><LF>
- give current ambient conditions state
Response:
EVG_0_OK<CR><LF>
- current ambient conditions state: unstable
Set Filter
Format:
FIS_n <CR><LF>
Response options:
FIS_OK <CR><LF>
- command carried out
FIS_E <CR><LF>
- error in-course of command execution, no parameter or incorrect
format
FIS_I <CR><LF>
- command understood but not accessible at this moment
n
- parameter, decimal value determining filter number
n
→
1 – very fast
2 – fast
3 – average
4 – slow
5 – very slow
Caution:
The numbering is assigned to a particular filter name and it is identical for all balance types.
Содержание PM 10.C32
Страница 1: ...C32 Series Balances Precision balances of PM C32 series IMMU 23 04 11 17 EN www radwag com USER MANUAL...
Страница 7: ...7 1 GENERAL INFORMATION 1 1 Dimensions PM xx C32 d 0 01g PM xx C32 d 0 1g...
Страница 8: ...8 PM xx C32 d 0 5g d 1g Indicator...
Страница 125: ...125...
Страница 128: ...128...
Страница 129: ...129...