ZENITH 140
INSTRUCTION MANUAL
132
Bargraph
P829 Bargraph
By default, the bar graph will be representative of the reading obtained, as
determined by the
Mode P100
. When
P100
=
4 (Average)
or
5
(Differential)
the bar graph can be assigned to be representative of the
level
on any of the two points of measurement. This parameter is
automatically set to the correct default option when selecting the
Mode
P100
and
Xducer
(
P101
), and under normal circumstances will not require
changing. The options, dependant on the
value
entered for
Mode P100
,
are as follows:
OPTION
DESCRIPTION
1 = Auxiliary (Optional)
Bargraph
will be representative of levels obtained
from optional
Aux. Input
.
2 = Xducer 1 (Default)
Bargraph
will be representative of levels obtained
from
Xducer 1
3 = Xducer 2
Bargraph
will be representative of levels obtained
from
Xducer 2
4 = Avg. Level or Diff
Bargraph
will be representative of levels obtained
from optional
Xducer 1