PULSAR MEASUREMENT
65
OPTION
DESCRIPTION
1 = Liquid (Default)
Used for liquids and flat solid materials.
2 = Solid
Used for sold material that is heaped or at an angle.
3 = Closed Tank
Use for applications within a closed tank or where a
secondary echo response may become focused to
create a larger echo than the first.
P103 Input 2
The 4-20 mA input can be used to replace the transducer for applications
where an ultrasonic transducer cannot be used.
Use this parameter to set the second input device when using in Average or
Differential Mode, and P101 Transducer = 0 (Auxiliary Input).
OPTION
DESCRIPTION
0
=
None (Default)
Use the optional mA input (Please consult Pulsar for
availability).
1 = dB3
Transducer is a dB3. Range = 0 to 3 metres
2 = dB6
Transducer is a dB6. Range = 0.3 to 6 metres
3 = dB10
Transducer is a dB10. Range = 0.3 to 10 metres
4 = dB15
Transducer is a dB15. Range = 0.5 to 15 metres
5 = dB25
Transducer is a dB25. Range = 0.6 to 25 metres
6 = dB40
Transducer is a dB40. Range = 1.2 to 40 metres
7 = dBS6
Transducer is a dBS6. Range = 0.2 to 6 metres
8 = dBMach3
Transducer is a dBMach3. Range 0.3 to 2.425 metres
10 = dB50
Transducer is a dB50. Range = 2 to 50 metres
*11 = dBR16
Transducer is a mmWave Radar. Range = 0.077 to 16 metres
*12 = dBR8
Transducer is a mmWave Radar. Range = 0.077 to 8 metres
*The signal emanates from the curved face of the Radar, but for the
purposes of measurement it is taken from the drip shield.
Содержание Zenith 140
Страница 1: ...Zenith 140 Instruction Manual ...
Страница 2: ...2 ...
Страница 14: ...ZENITH 140 INSTRUCTION MANUAL 14 EU Certificate of Conformity ...
Страница 17: ...PULSAR MEASUREMENT 17 Dimensions Fascia Mount The dimensions of the wall fixing holes are as shown below ...
Страница 19: ...PULSAR MEASUREMENT 19 Terminal connection details Fascia Mount The terminal details are as illustrated below ...
Страница 160: ...ZENITH 140 INSTRUCTION MANUAL 160 NOTES ...