ZENITH 140
INSTRUCTION MANUAL
34
Using the Serial Interface
The RS232 serial interface is used to communicate between the Zenith and a
PC using the optional Ultra PC and other associated Pulsar software
packages, to obtain information such as data logging and view echo traces
upload, download and save parameter files. In addition, it can also be used
to control or obtain information using a standard PC or other computer base
equipment. To do so, the settings for control are as follows:
baud rate
19,200
,
8 data bits
,
no parity
,
1 stop bit
.
The device should be connected as shown in
To use the device remotely, you need to
log on
to start, and
log off
when
finished. When
logged on
, the Zenith
will show ‘
Remote ON
’ on the display,
and “
Communicator OFF
” when
logged off
.
All commands should be followed by a carriage return
. When logged on, the
unit will respond either OK (or a value) if the command is accepted, or NO if
it is not.
To log on, send the command
/ACCESS:pppp where pppp is the passcode (P922).
To log off, send the command
/ACCESS:OFF
To read a parameter value, send the command
/Pxxx where xxx is the parameter you wish to read, and the Zenith 140 will
respond with the parameter value.
To set a parameter, send the command
/Pxxx:yy where xxx is the parameter number, and yy is the value you wish to
set it to.
Содержание Zenith 140
Страница 1: ...Zenith 140 Instruction Manual ...
Страница 2: ...2 ...
Страница 14: ...ZENITH 140 INSTRUCTION MANUAL 14 EU Certificate of Conformity ...
Страница 17: ...PULSAR MEASUREMENT 17 Dimensions Fascia Mount The dimensions of the wall fixing holes are as shown below ...
Страница 19: ...PULSAR MEASUREMENT 19 Terminal connection details Fascia Mount The terminal details are as illustrated below ...
Страница 160: ...ZENITH 140 INSTRUCTION MANUAL 160 NOTES ...