PT113
Instruction Manual
27
Parameter
Description
Secondary DNS :
Obtain secondary DNS manually.
Default is ‘208.67.220.220’.
If DHCP is selected, the DNS address should be automatically set.
Password
: Enter the password here to save changes or to change the password.
Default password:123456
6.3.4.4 PT113CO
The Gateway tab in AzComT is used only for setting a CANopen rack address between 1 and 127.
Default is '01'.
6.3.5 Back up Parameters and Calibration
Warning:
The previous parameter values and calibration coefficients in the PT113 will be lost after
loading a stored data back up into the instrument.
The parameter settings and calibration coefficients can be saved to a back up file after reading this data
from the PT113. This back up file can be restored to the PT113 after opening it with AzComT.
The back up feature provides a service advantage to PT113 transmitters. It is recommended to burn
these files to CD / DVD and keep this back up CD / DVD in the control cabinet and additionaly saving
the file in your PC.
6.4 PRE-COMMISSIONING TESTING
Scale performance testing should be performed before any field bus connection to external equipment.
These tests should include linearity, repeatability and eccentricity to ensure that the scale is weighing
correctly.
You can monitor testing with AzComT from the virtual display and/or status tab which has measurement
data, instrument data and software version. Taring, Zeroing and Clear keys are located in this tab for
easy testing of these functions.