PT113
Instruction Manual
35
Z
Zero (same as pressing the zero key)
Status Table:
A
Ack. Acknowledge, the command has been performed successfully
D
Dynamic, unstable weight
E
Errors except of H, L, O, +, −.
H
High voltage detected
I
The weight is in range
L
Low voltage detected
N
Nack. Not acknowledge, the command was not performed successfully
O
ADC out of range
S
Stable weight
X
Syntax error (the received command was not recognized)
+
Overload
−
Underload
Note: CHK, CR and LF will not be shown in the below data format descriptions in this section.
Commands and Responses:
A
|
Read all weight data
Command :
[ADR][A]
Response :
[ADR][A][STATUS][SIGN][NET W][SIGN][TARE W][SIGN][GROSS W]
Example :
Command : 01A
Response : 01AS+000123.4+000111.1+000234.5
01AD+000123.4+000111.1+000234.5
01AO (ADC out error)
Comments:
The response is net, tare and gross weight values or error status.
All weight data is transmitted immediately after receiving the command.
B
|
Read Gross weight
Command :
[ADR][B]
Response :
[ADR][B][STATUS][SIGN][WEIGHT VALUE]
Example :
Command : 01B
Response : 01BS+000123.4
(gross weight is stable and 123.4)
01BD+000123.4
(gross weight is dynamic and 123.4)
01B−
(under load)
Comments:
The response is the gross weight value (stable or dynamic) or error status.
Gross weight data is transmitted immediately after receiving command.
C
|
Clear the tare memory
Command :
[ADR][C]
Response :
[ADR][C][A]
(Memory cleared and the scale is in gross mode)
Comments:
The response status is always A in weighing or force mode.
D
|
Read count value immediately
Command :
[ADR][D]
Response :
[ADR][D][STATUS][SIGN][COUNT VALUE]
Example :
Command : 01D
Response : 01DD+00123400
01DO
(ADC out of range error)
01DX
(Not in count mode)
Comments:
The count value is sent immediately.