7 TROUBLE SHOOTING
7.1 ERRORS
The PT113 transmitters have been designed as very reliable and virtually error free instruments.
However if an error occurs, do not attempt to repair the equipment before you understand what caused
the error. Note the problem you are having and the status of the front panel, and try to find the problem
with the help of the table given below. Don’t let unauthorised people interfere with the instrument.
Error Code
Description
Actions to take / Possible cause
Err
ADC Error
− Re-energise the instrument.
− If seen again, replace the PT113.
Over Load
− Check the load, load cells and mountings.
− Check the load cell cables and connections
− Load cells or instrument could have a fault.
− Check the sense wires are connected.
Weight is too low
ADC Out
− Check the load, load cells and mountings.
− Check the load cell cables and connections
− Load cells or instrument could have a fault.
− Check the sense wires are connected.
System Error
− Re-energise the instrument.
− If seen again, replace the PT113.
High Voltage
Detected
− Check the power supply that the voltage is
within the required voltage range.
Low Voltage
Detected
− Check the power supply that the voltage is
within the required voltage range.
LED off LED on
LED Flashing
Table 5.5 - Error table
NOTE: By connecting to AzComT, additional information may be displayed.