PT113
Instruction Manual
58
Calibration Status (always 32 bit integer)
Dword of TxPDO 2 (T_DW2) descriptions when read command is ‘Calibration Status’.
Refer to RxPDO 3 (R_DW3) of ‘PLC Output to PT113CO Input’.
Bit Number
Low Dword of TxPDO 1 (T_UL1) Description
D31 … D16
Not in use
D15 … D8
Calibration
Errors
Bin
Dec
Descriptions
0000 0001
1
Calibration Timeout
- Restart calibration
0000 0010
2
ADC Error
- Re-energise the instrument
0000 0011
3
Instrument can not be calibrated
- Check load cell cable
- Re-energize the instrument
0010 0010
34
Instrument can not be calibrated
- Load cell signal is too low or too high
0010 0011
35
Calibration Error
- Calibration test weight is not enough
- Increase the test weight value. (Write test weight
value to PLC Output RxPDO 2 (R_DW2) to
PT113CO Input then restart the calibration)
- Check load cell connections
0010 0101
37
Scale unstable
- Wait until scale become stable
- Check grounding wiring
D7 ... D0
Calibration
Process
Status
0000 0001
1
System ready for calibration
0000 0011
3
Zero calibration in process
0000 0100
4
Span calibration in process
0000 1001
9
Error (Refer to Calibration Errors)
Table 12.11
- Calibration status
PLC Output to PT113CO Input
Bitwise order of a Dword:
Dword
(R/W)
D31 D30 D29 D28 D27 D26 D25 D24 D23 D22 D21 D20 D19 D18 D17 D16
D15 D14 D13 D12 D11 D10
D9
D8
D7
D6
D5
D4
D3
D2
D1
D0
TxPDO 1
(T_DW1)
Not in use
Not in use
TxPDO 2
(T_DW2)
The settings in RxPDO3 (R_DW3)
Define the contents for this Dword
TxPDO 3
(T_DW3)
Not in use
Not in use
Command List
Data Selection
New
Cmd