PT113
Instruction Manual
53
Calibration Status (always 32 bit integer)
2
nd
Dword descriptions are as below when the read command is ‘Calibration Status’. Refer to 3
rd
Dword
of PLC Output to PT113EI Input
Bit Number
2
nd
Dword Description
D31 … D16
Not in use
D15 … D8
Calibration
Errors
Bin
Dec
Descriptions
0000 0001
1
Calibration Time out
- Restart calibration
0000 0010
2
ADC Error
- Re-energize the instrument
0000 0011
3
Instrument can not be calibrated
- Check load cell cable
- Re-energise the instrument
0010 0010
34
Instrument can not be calibrated
- Load cell signal is very low or too high
0010 0011
35
Calibration Error
- Calibration loading is not enough
- Increase test weight loading (Write test weight
value to 2
nd
Dword of PLC Output to PT113EI
Input then restart the calibration)
- Check load cell connections
0010 0100
36
Calibration load value entry Error
- Test weight is too small. Increase the test weight
0010 0101
37
Scale unstable
- Wait until the scale becomes stable
- Check grounding wiring
D7... D0
Calibration
Process
Status
0000 0001
1
System ready for calibration
0000 0011
3
Zero calibration in process
0000 0100
4
Span calibration in process
0000 1001
9
Error (Refer to Calibration Errors 7.1ERRORS)