PT113
Instruction Manual
48
PLC Output to PT113PN Input
Bitwise order of a Dword:
Dword
(R/W)
D31 D30 D29 D28 D27 D26 D25 D24 D23 D22 D21 D20 D19 D18 D17 D16
D15 D14 D13 D12 D11 D10
D9
D8
D7
D6
D5
D4
D3
D2
D1
D0
1
st
Dword
Not in use
Not in use
2
nd
Dword
The settings in the 3
rd
Dword
Define the contents for this Dword
3
rd
Dword
Not in use
Not in use
Command List
Data Selection
New
Cmd
PLC Output to PT113PN Input 3
rd
Dword
Bit
Number
3
rd
Dword descriptions
31 … 11
Not in use
10 … 6
Command
List
00000
No command is activated
00001
Zero
00010
Tare
00011
Clear
00101
Adjust zero calibration
00110
Adjust span calibration (First load 2nd Dword with span test weight
value, then apply this command with New CMD)
01000
Operating mode selection (First
load 2nd Dword with selected
value, then apply this command
with New CMD)
Load 2nd Dword with one of these:
0: Count mode unipolar
1: Count mode bipolar
2: Force mode unipolar
3: Force mode bipolar
4: Weight mode unipolar
01001
Input range (mV) for Count
Mode
(1)
(First load 2nd Dword with
selected value, then apply this
command with New CMD)
Load 2nd Dword with one of these:
0: 5 mV (1mV/V)
1: 10 mV (2mV/V)
2: 15 mV (3mV/V)
3: 18 mV (3.6mV/V)
01010
Digital Filter
(First load 2nd Dword with
selected value, then apply this
command with New CMD)
Load 2nd Dword with one of these:
0: Fast
…
5: Medium
…
7: Default
…
9: Slow
5 … 1
Read Data
Selection
00000
Indicated weight
00001
Gross weight
00010
Tare weight
00011
Indicated weight (Floating point)
00100
Gross weight (Floating point)
00101
Tare weight (Floating weight)
10000
Calibration status
0
New CMD
Toggle
Apply commands which are listed in this table
(1) For bipolar input, the range is ± the value shown. i.e. 3 is ± 18mV