ALARMS, SAFETY, AND SELF-DIAGNOSIS
Tables 4–6 show the behaviour of the alarm (buzzer) and ring back function when pressing the BUZZER
STOP key.
Table 4. In cases other than those covered in Tables 5 and 6
Remote Alarm setting
Ring Back
setting
Buzzer from CO
2
incubator
Remote Alarm
When pressing
the BUZZER
STOP key
When the Ring Back
set time passes
When pressing
the BUZZER
STOP key
When the Ring Back
set time passes
ON: Remote alarm setting not connected with
BUZZER STOP key
ON
OFF
(Alarm is not
cancelled)
ON
ON
ON
(Under continuation)
OFF
OFF
OFF: Remote alarm setting connected with
BUZZER STOP key
ON
ON
OFF (Alarm is
not cancelled)
ON
OFF
OFF
OFF
Note:
Resolve the cause of the alarm, as the alarm itself is not deactivated by pressing the
BUZZER STOP
key.
Table 5. In cases of high-limit temperature alarm
Remote Alarm setting
Ring Back
setting
Buzzer from CO
2
incubator
Remote Alarm
When pressing
the BUZZER
STOP key
When the Ring Back
set time passes
When pressing
the BUZZER
STOP key
When the Ring Back
set time passes
ON: Remote alarm setting not connected with
BUZZER STOP key
ON
ON
ON
(Under continuation)
ON
(Continue)
ON
(Under continuation)
OFF
OFF: Remote alarm setting connected with
BUZZER STOP key
ON
OFF
Table 6. In cases of Error01, Error11, Error18, or door alarm*
1
Remote Alarm setting
Ring Back
setting
Buzzer from CO
2
incubator
Remote Alarm
When pressing
the BUZZER
STOP key
When the Ring Back
set time passes
When pressing
the BUZZER
STOP key
When the Ring Back
set time passes
ON: Remote alarm setting not connected with
BUZZER STOP key
ON
OFF
(Alarm is
cancelled)
OFF
(Alarm is already
cancelled)
OFF
(Alarm is
cancelled*
1
)
OFF
(Alarm is already
cancelled*
2
)
OFF
OFF: Remote alarm setting connected with
BUZZER STOP key
ON
OFF
*1:The remote alarm does not work for the door alarm.
Note:
When Err01 is activated, connect the new CO
2
gas cylinder and press the BUZZER STOP key to
stop the buzzer. In addition, when the optional MCO-21GC is installed and the gas supply is switched to
the reserve gas cylinder, press the BUZZER STOP key and replace the gas cylinder.