BASIC PARAMETERS
Setting CO
2
gas
Set the CO
2
gas control mode normal operation according to the following procedure. The incubator
automatically starts operation using these settings after power-on.
Setting values
:
ON
、
OFF
Factory setting : OFF
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display will change to the Menu screen.
2. When the cursor on the menu screen is on Set Value [Fig. 1], press
the enter key (ENTER).
▶
The display will change to the Set Value screen.
3. Move the cursor on the Set Value screen to CO
2
ON/OFF [Fig. 2]
using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The left side of the display will change to the CO
2
ON/OFF screen,
and the current set value (ON) will be displayed [Fig. 3].
4. Use the up/down keys (
△▽
) to change the control setting value.
Each time the up/down key is pressed the input value will change
between ON and OFF.
5. Press the enter key (ENTER).
▶
If it is set to ON: the input value will be confirmed and the display will
change to the CO
2
Setting screen.
If it is set to OFF: the input value will be confirmed and the display will
change to the Set Value screen.
6. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90
seconds has passed without any key operation.
<Right side of the display> [Fig. 3]
CO
2
ON/OFF
ON
ON / OFF
Menu
Set Value
Data Log
Lock
Alarms & Controls
<Left side of the display> [Fig. 1]
<Left side of the display> [Fig. 2]
Set Value
Temperature
:
37.0°C
CO
2
ON/OFF
: ON
CO
2
Setting
: 5.0%
Overtemp
: 53.0°C