53
ALARM PARAMETERS
Setting ring back delay
The function is that the alarm buzzer sounds again when the alarm state still continues after the set time of
ring back passed even though the alarm buzzer was stopped by pressing the BUZZER STOP key.
Settable range: 0 minute~99 minutes
Factory setting: 30 minutes
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display will change to the Menu screen.
2. Move the cursor on the Menu screen to Alarms & Controls [Fig. 1]
using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display will change to the Alarms & Controls screen.
3. Move the cursor on the Alarms & Controls screen to Other Alarm Set
[Fig. 2] using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display will change to the Other Alarm Set screen.
4. Move the cursor on the Other Alarm Set screen to Ring Back Delay
[Fig. 3] using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The right side of the display will change to the Ring Back Delay
screen, and the current setting value (30) will be displayed [Fig. 4].
5. Move the cursor using the left/right keys (
◁▷
) and select the digit you
want to change, and then use the up/down keys (
△▽
) to change the
number.
6. Press the enter key (ENTER).
▶
The input value will be confirmed and you will be returned to the
Other Alarm Set screen.
*If the entered setting value is out of the setting range, the following
screen will be displayed on the right side of the display: Press any key on
the control panel to change to the Ring Back Delay screen [Fig. 4].
7. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
Note: In the case of Err01 (CO
2
gas cylinder empty), Err11 (CO
2
sensor error) or Door alarm, the alarm is not
re-activated because the alarm itself is canceled by pressing the BUZZER STOP key. Incidentally in the case
of Err18 (UV lamp failure), the alarm is not re-activated if the BUZZER STOP key is pressed after the UV lamp
ON period elapses (refer to page 67~68).
<Left side of the display> [Fig. 1]
Menu
Set Value
Data Log
Lock
Alarms & Controls
Ring Back Delay
Invalid Value
Press Any Key
0min – 99min
<Right side of the display>
<Right side of the display> [Fig. 4]
Ring Back Delay
30
min
0min – 99min
<Left side of the display> [Fig. 2]
Alarms & Controls
Temp/Gas Alarm Set
Other Alarm Set
UV Setting
Gas Supply Setting
Date & Time
Keypad Click Set ▼
<Left side of the display> [Fig. 3]
Other Alarm Set
Door Alarm Delay
Ring Back Delay
Remote Alarm