29
HOME SCREEN
⑨
Outer door opened/closed status display
Outer door open: "Door:Open" is highlighted.
Outer door closed: "Door:Closed" will be
displayed.
⑩
UV lamp status display
For when using the optional UV lamp.
If the UV lamp is lit: "UV:ON" is highlighted.
If the UV lamp is off: "UV:OFF" will be displayed.
⑪
CO
2
gas supply line indicator A and B
When using the separately sold automatic CO
2
gas switching equipment, the CO
2
supplying CO
2
gas supply line (CO
2
gas pipe connection port)
currently in use is highlighted.
It blinks if the CO
2
cylinder is empty. (See pages
60~62 for reference)
⑫
Message count display
If the total number of messages is more than 1, it
will be displayed.
The display shows the "number of the message
currently displayed/total number of messages".
If the left/right keys (
◁▷
) are displayed, the
contents of the message number can be switched
using the left and right keys.
⑬
Error/Alarm display
Displays the currently occurring error/alarm. (See
pages 67~68)
If there is an error: "Err
〇〇
" (
〇〇
is the error
code) will be displayed
If it's an alarm: "Warning" will be displayed.
⑭
Message display
Errors/Alarms/Other messages are displayed
accordingly. (See pages 67~68)
Notes:
The display starts scrolling from right to left once
the unit is in the following four conditions.
• Not in an alarm condition (no error messages are
displayed).
• Temperature display and CO
2
density display
with no blinking.
• The outer door is closed.
• No keys being pressed over 30 seconds.
Scrolling can be stopped in either one of the
following conditions.
• In an alarm condition (error messages are
displayed).
• Temperature display or CO
2
density display with
blinking.
• The outer door is opened.
• Any key is pressed.
⑪
⑨
⑩
⑫
⑬
⑭
<Left side of the display>
<Right side of the display>