BASIC PARAMETERS
Setting high limit temperature alarm
The high limit temperature alarm is different from the Automatic set temperature alarm (page 48), and it is
independent temperature alarm. In case the chamber temperature exceeds the temperature of the high limit
temperature alarm, this alarm is activated.
Set the High Limit normal operation according to the following procedure. The incubator automatically starts
operation using these settings after power-on.
Settable range : 20.0
o
C~53.0
o
C
Factory setting : 53.0
o
C
Note:
·
Set the temperature of the high limit temperature alarm after the chamber temperature is stable at the set value.
·
Set the high limit temperature alarm to at least 1
o
C higher than the chamber set temperature.
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display will change to the Menu screen.
2. When the cursor on the menu screen is on Set Value [Fig. 1], press
the enter key (ENTER).
▶
The display will change to the Set Value screen.
3. Move the cursor on the Set Value screen to Overtemp [Fig. 2] using
the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The right side of the screen will change to the Overtemp screen, and
the current setting value (53.0) will be displayed [Fig.3].
4. Move the cursor using the left/right keys (
◁▷
) and select the digit you
want to change, and then use the up/down keys (
△▽
) to change the
number.
5. Press the enter key (ENTER).
▶
The input value will be confirmed and you will be returned to the Set
Value screen.
*If the entered setting value is out of the setting range, the following
screen will be displayed on the right side of the display: Press any key
on the control panel to change to the Overtemp screen [Fig. 3].
6. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
Overtemp
Invalid Value
Press Any Key
20.0°C – 53.0°C
<Right side of the display>
<Right side of the display> [Fig. 3]
Overtemp
53.0
°C
20.0°C – 53.0°C
<Left side of the display> [Fig. 1]
Menu
Set Value
Data Log
Lock
Alarms & Controls
<Left side of the display> [Fig. 2]
Set Value
Temperature
:
37.0°C
CO
2
ON/OFF
: ON
CO
2
Setting
: 5.0%
Overtemp
: 53.0°C