![Phcbi MCO-170AC Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/phcbi/mco-170ac/mco-170ac_operating-instructions-manual_1536737014.webp)
14
INCUBATOR COMPONENTS
Control panel
①
OLED (Organic Light Emitting Diode) display
The present chamber temperature/CO
2
density, menu screen or input screen, etc. are displayed.
②
POWER/ALARM indicator
In normal operation, the green LED lamp lights up, and during alarm condition, the red LED lamp blinks to
indicate alarm.
③
Up/down/left/right key (
△▽
◁▷
)
On the setting screen, pressing this key changes the display.
If the left / right key (
◁▷
) is displayed at the message of the home screen, you can switch the contents of
the message number with the left and right keys.
④
Menu key (MENU/HOME)
On the “home screen”: pressing this key leads the menu screen. It is possible to set various setting on the
menu screen.
On the screen “other than the home screen”: pressing this key leads the home screen.
⑤
Cancel key (CANCEL
)
On the setting screen, pressing this key changes into the screen of one level up.
⑥
Buzzer stop key (BUZZER STOP)
The buzzer is silenced by pressing this key when the buzzer sounds.
However, when the ring back is ON, the buzzer will sound again when the set time of ring back passed and
the alarm state still continues. Refer to page 53 and 67~69.
Note: It is not possible to silence the buzzer for the high limit temperature alarm.
⑦
Enter key (ENTER)
Press this to select the menu item. Press this key to enter the required value during the setting procedure.
⑧
USB port
Insert a USB memory device to export data logs.
Do not insert anything other than a USB memory device.
①
②
③
④
⑤
⑥
⑦
⑧