![Phcbi MCO-170AC Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/phcbi/mco-170ac/mco-170ac_operating-instructions-manual_1536737040.webp)
40
OPERATION/ALARM LOG
8. When outputting is complete, "Export complete" will be displayed [Fig.
6]. Press any key to return to the Data Log Export screen.
Note:
Even after the export of operation log data is complete,
operation log data saved in the incubator are not deleted.
*The following messages may come up on the right side of the display:
USB memory is disconnected: USB memory is not inserted
USB memory is full: there is no available space in the USB memory
An error occurred: failed to output to USB memory
No Data: there is no data to output
▶
Press any key on the control panel to return to the Data Log Export
screen.
9. Remove the USB memory from the USB port.
Reference:
・
A Log folder is created in the USB memory where the output file is saved in CSV format. The output file
name will be displayed with the device ID, date and time of output, and data name. The date format can be
changed in the settings (see page 63).
(Example) When the device is set to "A00001" and data from Jan. 1, 2019 - Oct. 1, 2019 is output using All:
YY/MM/DD
DD/MM/YY
The oldest date of
operation log - date of
that day
A00001_20190101-20191001_DataLog.csv
A00001_20190101-20191001_DoorLog.csv
A00001_01Jan2019-01Oct2019_DataLog.csv
A00001_01Jan2019-01Oct2019_DoorLog.csv
(Example) When the device is set to "A00001" and the data from Jan. 1, 2019 is output using 24 Hours:
YY/MM/DD
DD/MM/YY
Specified date
A00001_20190101_DataLog.csv
A00001_20190101_DoorLog.csv
A00001_01Jan2019_DataLog.csv
A00001_01Jan2019_DoorLog.csv
*If it is set to All. the outer door opening/closing history data file is output together.
*If the device ID is not set, the default setting “000000” will be displayed.
*If duplicate file names are used, the data that is output later will be overwritten.
10. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
<Right side of the display> [Fig. 6]
Export complete