BASIC PARAMETERS
Setting temperature
Set the chamber temperature normal operation according to the following procedure. The incubator
automatically starts operation using these settings after power-on.
Settable range : 0.0
o
C
〜
50.0
o
C
Factory setting : 37.0
o
C
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display will change to the Menu screen.
2. When the cursor on the menu screen is on Set Value [Fig. 1], press
the enter key (ENTER).
▶
The display will change to the Set Value screen.
3. When the cursor on the Set Value screen is on Temperature [Fig. 2],
press the enter key (ENTER).
▶
The right side of the display will change to the Temperature screen,
and the current set value (37.0) will be displayed [Fig. 3].
4. Move the cursor using the left/right keys (
◁▷
) and select the digit you
want to change, and then use the up/down keys (
△▽
) to change the
number.
5. Press the enter key (ENTER).
▶
The input value will be confirmed and you will be returned to the Set
Value screen.
*If the entered setting value is out of the setting range, the following
screen will be displayed on the right side of the display: Press any key
on the control panel to change to the Temperature screen
[Fig. 3].
6. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
<Right side of the display>
Temperature
Invalid Value
Press Any Key
0.0°C – 50.0°C
<Left side of the display> [Fig. 2]
Set Value
Temperature
:
37.0°C
CO
2
ON/OFF
: ON
CO
2
Setting
: 5.0%
Overtemp
: 53.0°C
<Left side of the display> [Fig. 1]
Menu
Set Value
Data Log
Lock
Alarms & Controls
<Right side of the
display> [Fig. 3]
Temperature
37.0
°C
0.0°C – 50.0°C