Section 6-1-4
6-1-4 Maintenance mode #3: Offset adjustment
In this mode, the offset correction values for main/sub scanning are automatically calculated to satisfy the following offset
values:
Offset value
Main scanning: The largest offset of A6 or larger size of document shall be: ± 24 dot
Sub-scanning: The largest offset of A6 or larger size of document shall be: ± 33 dot
Before this adjustment, please prepare the Test sheet described in Figure 6-1-3.
Also, before adjusting offset (ADF front or back, or FB), run Main-scanning magnification adjustment (ADF front or back, or
FB) by referring to Section 6-1-3.
[How to start]
(1) From screen T04, Press the Function (
△
or
▽
) button to select (Maintenance mode #3) and press the Scan/Stop button.
A number is shown on the Function No. Display indicating the location of the offset to be adjusted.
Function No.
Display
Offset to be adjusted
Remarks
0 ADF
front
Default
1 ADF
back
2
FB
FB models (fi-6240/fi-6230) only
(2) Change the selection by pressing Function button.
(3) For the ADF adjustment, set the test sheet (see Figure 6-1-3) on the ADF paper chute (Chute Unit), and adjust the side guide
to the width of the test sheet.
For the FB adjustment, set the test sheet (see Figure 6-1-3) on the Document bed to be aligned with the reference point (left of
the Document bed, front edge), and remove the Document Cover. Avoid working near light source such as fluorescent lamp.
(4) Press Scan/Stop button to start the adjustment operation.
[How to end]
Press Send to button during operation. The operation stops and the Maintenance mode selection screen (T04) appears.