08
July 27, 09 K.Okada
A.Miyoshi I.Fujioka
Refer to Revision Record on page 2.
TITLE
fi-6140/fi-6240/fi-6130/fi-6230/fi-614PR
Maintenance Manual
07
Nov.12, 08 K.Okada
T.Anzai
I.Fujioka
Refer to Revision Record on page 2.
CUST.
06
Mar.10, 08 K.Okada
T.Anzai
I.Fujioka
Refer to Revision Record on page 2.
DRAW
No.
P1PA03540-B0XX/6
Rev. DATE DESIGN CHECK APPR. DESCRIPTION
Design
July 27, 2007
K.Okada
CHECK
K.Okada
APPR.
T.Anzai
PFU LIMITED
Page
162
/
257
Section 6-1-3
(2) When the magnification adjustment is terminated abnormally
Screen T25
Function No.
Display
Scanner status
Displays “c” without blinking.
.
Note: The major reason of abnormal termination is incorrect setting of the test sheet. Set the test sheet correctly and try the
magnification adjustment again.
<Available buttons at screen T25>
Function
button: Displays error information (screen T26)
Send to button: Terminates this mode and returns to screen T04.
Screen T26
Adjustment mode (*1)
Function
No.
Display
Description
0
1,2
3
4
Countermeasure when abnormal
termination frequently occurs
1: cannot detect the leading
edge of the document
√
√
2: cannot detect the left edge of
the document
√
√
4: cannot detect the right edge
of the document
√
√
5: Excessive skew A
√
√
6: cannot detect the trailing
edge of the document
√
√
7: Excessive skew B
√
√
Conduct necessary operation by referring
to step (2) and later in Section 4-3-7 or
Section 4-3-10.
(*1) 0 : ADF sub-scanning magnification adjustment
1,2: ADF front/back main scanning magnification adjustment
3: FB sub-scanning magnification adjustment
4: FB main scanning magnification adjustment
Skew A and B are calculated as follows:
Skew A = a – b
Skew B = c – d
<Available button at screen T26>
Send to button: Terminates this mode and returns to screen T04.
a
b
c
d
1
2
5
6
7
4
3