257
Section 5-10-4
5-10-4 OPT Unit (for backside scanning)
- Refer to Section 3-6 for the part number of the replacement part.
- Refer to Appendix for the specification of the screws.
<Removal>
1. Remove Chuter Unit.
(1) Referring to Section 5-8-1 “Chuter Unit (ADF Paper Chute),” remove the Chuter Unit.
2. Remove Panel PCA cable.
(2) Referring to steps of Section 5-10-9 “Panel PCA,” disconnect the Panel PCA cable.
Note: There is no need to save EEPROM data.
3. Remove Top Cover.
(3) Referring to steps (3) ~ (5) of Section 5-10-11 “Top Cover,” remove the Top Cover.
4. Remove OPT Unit.
(4) - Remove the CCD cable. 1
Note: Pinching the root of connector of the CCD board to support, pull out the CCD cable straight.
- Remove the screw A (circle below) on the OPT Hold Plate, and then remove the OPT Hold Plate. 2
- Lift up the OPT Unit 3 to remove.
OPT Unit
2
3
1