NECTOR
Page
30
OPERATION AND MAINTENANCE MANUAL
Rev. 03-21
-
pg3:
Export data in secure format compatible with the TeleNET supervision software.
-
CSv:
Export data in standard table text format.
Confirm the save with the key
.
5.
Wait:
a short beep will sound when the save is complete.
6. In the event of an error relating to the USB memory, a long sound signal is emitted and the
flashing wording
Eu
is displayed with one of the following error codes:
1
– disconnect during save or offline memory
2
– Physical error / unable to write to disk
3
– invalid path
4
– access prohibited
5
- read-only unit
6
– incorrect file system/invalid device name
7
– 999 files (pg3 or csv) on USB exceeded
8
– generic USB alarm
9
– parameter import error
7. In the event of an error while saving data, it will be necessary to remove the cause and to
repeat the operation.
8. After saving, remove the USB stick from the panel and insert it into your computer.
9.
Use TeleNET's “Automatic Import” function to simply import data in “pg3” format, or display
“CSv” data via a spreadsheet.
Refer to the TeleNET manual for a better understanding of the functions and options available
including the importing of data, the viewing of recordings and alarms, customizable graphics,
unique tool identification.
Note:
The file names *.pg3 and *.csv contain the instrument serial number. In order to allow
TeleNET to import the data correctly, the names of the exported files should not be changed.