50
GN Otometrics A/S
Tests
AccuScreen PRO User Manual
If
ABR
wins, the result is a
PASS
.
Test duration for babies:
Approximately 1½ minutes on babies with normal hearing.
•
Add comments and view results as required. See
‘Interpretation of the results” on page 54
•
You can now scroll to select the patient's other ear on the
screen and, if required, reposition the baby so that you can test
the other ear.
•
Before you continue testing the other ear, check to make sure
that the probe is clean. See also
6.2
Monitoring the test
6.2.1
TEOAE tests
The LED flashes green during the test.
If the LED is red, this indicates an error during testing:
•
The probe is unstable, or
•
Artifacts are present, indicating high environmental noise or a
noisy baby.
In these cases, check the test conditions and make sure that the probe
fits.
If required, press
STOP
to stop the test.
Tests with a
Stopped
result are not saved. Accordingly, they cannot be
retrieved in the
View results
screens and they cannot be printed
out.
If the message
Unstable
appears please check if the probe fits correct-
ly.
If you press
OPT
, the artifact and stimulus values are shown on the
display:
Содержание Madsen AccuScreen PRO
Страница 8: ...User Manual AccuScreen PRO 8 GN Otometrics A S ...
Страница 12: ...12 GN Otometrics A S Introduction AccuScreen PRO User Manual ...
Страница 20: ...20 GN Otometrics A S Test Methods AccuScreen PRO User Manual ...
Страница 26: ...26 GN Otometrics A S The AccuScreen PRO AccuScreen PRO User Manual ...
Страница 32: ...32 GN Otometrics A S Before You Start AccuScreen PRO User Manual ...
Страница 58: ...58 GN Otometrics A S Tests AccuScreen PRO User Manual ...
Страница 60: ...60 GN Otometrics A S Quick Test AccuScreen PRO User Manual ...
Страница 70: ...70 GN Otometrics A S The Transfer Delete Data Menu AccuScreen PRO User Manual ...
Страница 86: ...86 GN Otometrics A S Maintenance AccuScreen PRO User Manual ...
Страница 98: ...98 GN Otometrics A S User Manual AccuScreen PRO ...