92
GN Otometrics A/S
Technical Specifications
AccuScreen PRO User Manual
15.3
Probe
15.4
Measurement techniques
Charge termination (HL to EL):
-
Δ
U: approx. 0.7% with 5 cells (approx. 12 mV/cell).
The -
Δ
U monitoring is activated after approx. 7 min. following
commencement of charging
Maximum charging time:
2½ hours ±12%
Charging control:
The cell voltage is measured and detected every 107 sec. during
a zero current of 80 ms.
Probe plug:
12-pin
Probe cable:
Flexible, shielded cable, approx. length: 1 m (3’)
Frequency response:
200 Hz to 4.2 kHz (measured in 1.5 cm
3
test cavity)
Speakers:
Type:
Magnetic: ED 1913/Knowles low distortion
Signal bandwidth:
1.7 to 4.8 kHz
Nominal impedance (at 1 kHz):
1500
Ω
Nominal DC resistance:
375
Ω
Microphone:
Type:
Electret, EM 3046/Knowles, with integrated FET preamplifier
Signal bandwidth:
Front-end sensitivity:
1.7 to 6.5 kHz
Approx. 55 dB below 1V/microbar (0.1 N/m
2
)
Output impedance:
2800 to 6800
Ω
(4400
Ω
nominal)
Operating voltage:
0.9 to 20 V DC
Dimensions:
Probe body:
Probe tip:
10 mm Ø x 35 mm (0.4” Ø x 1.4”)
3.5 mm Ø x 10 mm (0.14” Ø x 0.4”)
Probe adapter:
4.0 mm to 12 mm Ø (0.16” to 0.47”) silicone
Weight:
Probe incl. probe adapter:
4 grams
Probe tip (TEOAE):
Exchangeable, green
Probe tip (DPOAE):
Exchangeable, blue
Eartips:
Standard (cylindric):
Mushroom:
Tree tip:
8 sizes (4 - 12 mm)
1 size (6.5 mm)
1 size (4 - 7 mm)
TEOAE:
Sampling rate:
16 kHz
Evaluation method:
AccuScreen binomial statistics
Stimulus:
Non-linear click sequence
Stimulus level:
70-84dB SPL (45-60 dB HL, self calibration depending on ear
canal volume
Click rate:
Approx. 60 Hz
Frequency range:
1.4 to 4 kHz
Display:
Statistical waveform, measurement progress, TEOAE detection
level, noise level
DPOAE:
Sampling rate:
12.8 kHz
Evaluation method:
Phase statistics
Stimulus:
Primary tone pair, f2/f1 = 1.24
Frame length:
512 samples
Test frequencies (screening):
f2 = 2, 2.5, 3.2, 4 kHz
Test frequencies (diagnostic/mul-
tilevel):
f2 = 1.5, 2, 3, 4 kHz
Test level (screening):
l1/l2 = 59/50 dB SPL
Содержание Madsen AccuScreen PRO
Страница 8: ...User Manual AccuScreen PRO 8 GN Otometrics A S ...
Страница 12: ...12 GN Otometrics A S Introduction AccuScreen PRO User Manual ...
Страница 20: ...20 GN Otometrics A S Test Methods AccuScreen PRO User Manual ...
Страница 26: ...26 GN Otometrics A S The AccuScreen PRO AccuScreen PRO User Manual ...
Страница 32: ...32 GN Otometrics A S Before You Start AccuScreen PRO User Manual ...
Страница 58: ...58 GN Otometrics A S Tests AccuScreen PRO User Manual ...
Страница 60: ...60 GN Otometrics A S Quick Test AccuScreen PRO User Manual ...
Страница 70: ...70 GN Otometrics A S The Transfer Delete Data Menu AccuScreen PRO User Manual ...
Страница 86: ...86 GN Otometrics A S Maintenance AccuScreen PRO User Manual ...
Страница 98: ...98 GN Otometrics A S User Manual AccuScreen PRO ...