90
GN Otometrics A/S
Status and error Messages
AccuScreen PRO User Manual
Poor testing condi-
tions!
• When there is too much noise in
more than 75% of testing time,
EEG and/or interference is too
high.
• Improve testing conditions.
• Reduce noise.
• Check interference from electri-
cal installations.
• Calm baby
Printer error
• -
• Check label printer manual.
Probe Error!
Calibration could not be finished
properly.
• Refit probe.
• Clean the probe tip.
• Check acoustic filter.
• Perform a probe test.
Ringing! (during
TEOAE test only)
• Poor probe fit.
• Reposition probe.
Stimulus too high
(duing ABR test only)
• ear probe slipped during ABR
measure
• Refit probe and restart test
Stimulus too high!
• Probefit not good.
• Probe not in ear.
• Reposition probe.
System Error
• System malfunction.
• Contact your service depart-
ment.
Testing not possible!
• Impedance of electrodes or bal-
ance is too high. EEG values are
constantly high.
• Check electrodes and replace
them if necessary. Repeat test if
the patient is quiet.
Testing not possible!
Please disconnect the
label printer!
• Testing cannot be done while the
printer is connected to the
device.
• Disconnect the label printer
from the device.
The calibration date
has expired. Please
have the device cali-
brated soon.
• The calibration date has expired.
• Contact your service depart-
ment for calibration.
Too much noise!
• Baby or testing environment is
too noisy.
• Reduce noise.
• Try to calm baby or wait until
baby is calm.
Unstable! (during
TEOAE test only)
• Probe changes position, e.g.
when baby moves its head.
• Probe not in ear.
• Reposition probe when baby is
calm.
• Reposition probe in ear.
Wrong probe
• A TEOAE probe is connected for
DPOAE testing.
• Connect DPOAE probe instead.
Error message
Cause
Solution
Содержание Madsen AccuScreen PRO
Страница 8: ...User Manual AccuScreen PRO 8 GN Otometrics A S ...
Страница 12: ...12 GN Otometrics A S Introduction AccuScreen PRO User Manual ...
Страница 20: ...20 GN Otometrics A S Test Methods AccuScreen PRO User Manual ...
Страница 26: ...26 GN Otometrics A S The AccuScreen PRO AccuScreen PRO User Manual ...
Страница 32: ...32 GN Otometrics A S Before You Start AccuScreen PRO User Manual ...
Страница 58: ...58 GN Otometrics A S Tests AccuScreen PRO User Manual ...
Страница 60: ...60 GN Otometrics A S Quick Test AccuScreen PRO User Manual ...
Страница 70: ...70 GN Otometrics A S The Transfer Delete Data Menu AccuScreen PRO User Manual ...
Страница 86: ...86 GN Otometrics A S Maintenance AccuScreen PRO User Manual ...
Страница 98: ...98 GN Otometrics A S User Manual AccuScreen PRO ...