72
GN Otometrics A/S
Options
AccuScreen PRO User Manual
11.1.1
Probe test
1.
Select
Probe test
to check that the probe functions correctly.
You will get information about:
•
the probe type
•
the serial number of the probe
•
the date of the last calibration
2.
Insert the probe tip without eartip in the test cavity, and press
any key to start the probe test.
If the probe functions correctly, the message "Probe OK!"
appears.
If the message "Probe failed" appears, press any key to see a
list of possible errors.
3.
Press any key to return to the
SELF TEST
menu.