54
GN Otometrics A/S
Tests
AccuScreen PRO User Manual
A list of all stored results for the selected patient is shown. See
also
Section 6.4, ‘Interpretation of the results” on page 54
The black bar indicates the most recent test result or comment.
2.
Press
OK
to view.
6.4
Interpretation of the results
A patient screened by the AccuScreen PRO is either given a
PASS
or
a
REFER
.
If the result is a
PASS
, no further action is required.
If the result is a
REFER
, it is recommended that you retest the pa-
tient. If the patient is given yet another
REFER
, this patient should
be referred for further testing.
The results screen is explained in the following:
PS
=
PASS
Testing successful (OAE or ABR present)
RE
=
REFER
OAE or ABR criterion not reached
* Date/Time format depends on the setting in
. The year of the test is stored as well and
can be seen if the test is transferred to AccuLink.
6.4.1
TEOAE
Applies to AccuScreen PRO T, TA, TD, TDA
Test result PASS
A
PASS
is a
PASS
, regardless of the
A
and
S
levels (see
Test type:
TE=TEOAE
DP=DPOAE
DD=DPOAE Diagnostic
AA=Automated ABR
CO=Comment
Date*
Result
Right or Left ear
Test time
Comments
Содержание Madsen AccuScreen PRO
Страница 8: ...User Manual AccuScreen PRO 8 GN Otometrics A S ...
Страница 12: ...12 GN Otometrics A S Introduction AccuScreen PRO User Manual ...
Страница 20: ...20 GN Otometrics A S Test Methods AccuScreen PRO User Manual ...
Страница 26: ...26 GN Otometrics A S The AccuScreen PRO AccuScreen PRO User Manual ...
Страница 32: ...32 GN Otometrics A S Before You Start AccuScreen PRO User Manual ...
Страница 58: ...58 GN Otometrics A S Tests AccuScreen PRO User Manual ...
Страница 60: ...60 GN Otometrics A S Quick Test AccuScreen PRO User Manual ...
Страница 70: ...70 GN Otometrics A S The Transfer Delete Data Menu AccuScreen PRO User Manual ...
Страница 86: ...86 GN Otometrics A S Maintenance AccuScreen PRO User Manual ...
Страница 98: ...98 GN Otometrics A S User Manual AccuScreen PRO ...