GN Otometrics A/S
Tests
AccuScreen PRO User Manual
The keynote (
) on the display indicates which frequency is cur-
rently being tested.
The results are indicated by the following symbols:
9
PASS
R
REFER
S
Skipped/stopped
If the result is a
PASS
, a
PASS
message appears before the detailed
results are shown.
At the end of each test a result display appears automatically, show-
ing the individual results of the tested frequencies.
The results are ordered and shown in increasing frequencies:
2000 Hz
2500 Hz
3200 Hz
4000 Hz
Only L2 is shown. The L1/L2 pair is 59/50 dB SPL.
The complete test is terminated successfully with a
PASS
if DPOAEs
could be registered in at least three frequency ranges/areas.
•
Add comments and view results as required. See
Section 6.4, ‘Interpretation of
•
You can now scroll to select the patient's other ear on the
screen and, if required, reposition the patient so that you can
test the other ear.
•
Before you continue testing the other ear, check to make sure
that the probe is clean. See also
PASS
Содержание Madsen AccuScreen PRO
Страница 8: ...User Manual AccuScreen PRO 8 GN Otometrics A S ...
Страница 12: ...12 GN Otometrics A S Introduction AccuScreen PRO User Manual ...
Страница 20: ...20 GN Otometrics A S Test Methods AccuScreen PRO User Manual ...
Страница 26: ...26 GN Otometrics A S The AccuScreen PRO AccuScreen PRO User Manual ...
Страница 32: ...32 GN Otometrics A S Before You Start AccuScreen PRO User Manual ...
Страница 58: ...58 GN Otometrics A S Tests AccuScreen PRO User Manual ...
Страница 60: ...60 GN Otometrics A S Quick Test AccuScreen PRO User Manual ...
Страница 70: ...70 GN Otometrics A S The Transfer Delete Data Menu AccuScreen PRO User Manual ...
Страница 86: ...86 GN Otometrics A S Maintenance AccuScreen PRO User Manual ...
Страница 98: ...98 GN Otometrics A S User Manual AccuScreen PRO ...